Hot: Skin Analyzer PDT Skin Machine Hair Growth Device Home Phototherapy
A phototherapy device is specified in milliwatts per square centimetre and prescribed in seconds. Between those two numbers sits everything that can quietly go wrong: an ageing lamp, a dusty reflector, a lamp replaced from a different batch, a patient standing ten centimetres closer.
Dose accuracy is not a laboratory concern. It is the difference between a protocol that works and one that burns a patient in week three.

A device that logs dose turns calibration into a reading rather than an argument.
A calibrated radiometer measures irradiance at the treatment plane. Compare that reading with the value on the device label and with the previous reading for the same unit. A fall of more than about 20 percent, or any reading outside manufacturer tolerance, means recalibrated dose tables or lamp replacement.
A workable routine is monthly output checks on high-use clinic units, quarterly on low-use units, and a full service with lamp inspection each year. Every reading should be signed and dated, and the meter itself needs an annual calibration certificate from a traceable laboratory.

Small units drift fastest because the lamp runs close to its thermal limit.
A device sold without a calibration path becomes a support problem. Supply a radiometer option, a one-page measurement procedure and a device log page, and offer a recalibration service or a local partner who can perform it. Clinics buy dose accuracy from whoever can demonstrate it.
Ask for our calibration log template, radiometer specification sheet and lamp replacement schedule.
【Editor】admin
Address
2nd Floor, 1-101,Building 6, Xuzhou , LIANDO U Valley, 6# Leye Road, Damiao Street, Jiangsu Province, China
Email
info@diosole.com
Tel/Whatsapp
+86 18914846950
Facebook
18914846950